Does the multi-lane mouse treadmill count air puffs toward the Warn/Alert limit?

Explains why the multi-lane mouse treadmill counts only shocks, not air puffs, and how the Warn/Alert limit works as a result.

Updated Nov 8, 2026
2 min read

No. The multi-lane mouse treadmill does not count, display or record air puffs. This is how the treadmill is designed. It is not a fault, and it does not depend on which control box you use. A replacement control box works the same way.

Because air puffs are not counted, the Warn/Alert limit applies only to shocks:

  • Air puff only: No stimulus count is kept, so the Warn/Alert limit does not stop a lane based on air puffs.
  • Air puff and shock both on: Only shocks are counted and reported. When a lane reaches the shock limit, time and distance stop for that lane.
  • Shock only: Shocks are counted, and the Warn/Alert limit works as expected.

Using the Warn/Alert limit

  1. Turn on the shock stimulus for the lanes you want to limit.
  2. Set the Warn/Alert limit to the number of shocks you want to allow.
  3. Start the run.
  4. Watch the shock count for each lane. When a lane reaches the limit, its time and distance stop.

If your protocol needs a per-lane air puff count, record air puffs by another method during the session. The treadmill will not include them in its data.

If you have questions about stimulus counting or your protocol, write to [email protected] or call +1 847 983 3672.